LABORATORY MEASUREMENTS - Laboratory Temperature
Basic types of calibrated instruments:
- Glass thermometers and electrical *
- instruments for measuring and recording temperature
- temperature and humidity sensors
- instruments for measuring and recording relative humidity and air temperature
- thermostats liquid *
- thermal and climatic chamber *
* Possibility of making a calibration with the customer.
|
|
|
The scope of accreditation for the laboratory temperature.
Calibration in accordance with the requirements of PN-EN ISO / IEC 17025:2005 certified by Polskie Centrum Akredytacji.
|
| Name of physical size and type of measuring instrument |
Meas |
Zdolność pomiarowa CMC |
Cat. Lab. |
Remarks |
|
| 19 Temperatura |
|
|
|
|
| 19.01N Termometria elektryczna |
|
|
|
|
| Electrical and electronic thermometers |
(-30 ÷ 105) °C (105 ÷ 200) °C (-20 ÷ 120) °C |
0,1 °C 0,16 °C 0,3 °C |
S S P |
|
| Indicators of temperature (temperature meters) |
(-200 ÷ 1820) °C (-200 ÷ 850) °C |
0,8 °C 1) 0,2 °C 2) |
S
|
|
| Thermocouples and thermoelectric sensors |
(0 ÷ 1100) °C |
0,9 °C |
S
|
|
| Resistance thermometer sensors |
(-30 ÷ 105) °C (105 ÷ 200) °C 0 °C |
0,08 °C 0,16 °C 0,05 °C |
S
|
|
| Thermostatic chamber and climatic conditions |
(-30 ÷ 200) °C (-30 ÷ 200) °C |
0,3 °C 3) 4) 0,3 °C 3) 4) |
S P |
|
| Thermostats liquid |
(-30 ÷ 200) °C (-30 ÷ 200) °C |
0,3 °C 3) 4) 0,3 °C 3) 4) |
S P |
|
| 19.02N Termometria nieelektryczna |
|
|
|
|
| liquid glass thermometers |
(-30 ÷ 105) °C (105 ÷ 200) °C |
0,08 °C 0,16 °C |
S
|
graduated to not less than 0,1 °C |
| Manometric thermometers |
(0 ÷ 200) °C |
0,3 °C |
S
|
|
14 Humidity
14.01 humidity |
|
|
|
Calibration in the range (0÷40) °C |
| Hygrometers and termohigrometry for relative humidity and air temperature |
(40 ÷ 90) %RH
(30 ÷ 90) %RH
(20 ÷ 90) %RH
(20 ÷ 90) %RH
(20 ÷ 80) %RH
(0 ÷ 40) ºC
|
(2,0 ÷ 2,9) %RH
(1,3 ÷ 1,8) %RH
(1,1 ÷ 1,8) %RH
(1,1 ÷ 1,8) %RH
(1,1 ÷ 1,5) %RH
0,25 ºC
|
S
|
dla 0 °C
dla 10 °C
dla 20 °C
dla 30 °C
dla 40 °C
|
*) Expanded uncertainty at a confidence level of approximately 95%.
1) Calibration using a standard indirect thermometric characteristics of thermocouples specified in BS EN 60584-1:1997 Thermocouples - Characteristics.
2) Indirect calibration using standard thermometric sensor characteristics of industrial platinum resistance thermometers, given in BS EN 60751 + A2: 1997 Sensors for industrial platinum resistance thermometers.
3) The ability to measure the simultaneous recording of 30 points in the space of the chamber.
4) The value of the best possible measurement concerns a single measuring point in space.
|
|
Outside the scope of accreditation the calibration includes:
- temperature measurement circuits
- temperature transducers
- pyrometers
- distributions of temperature and humidity in rooms
The feasibility of off-site calibration laboratory
Calibration based on procedures in accordance with the requirements of ISO 9001, as evidenced by a certificate of Polski Rejestr Statków.
|
Detailed information can be obtained by calling the following number:
| Head Office |
tel. 046 837 04 44, fax. 046 837 01 00 |
|
|
LABORATORY MANAGER Sławomir Wróbel |
tel./fax 046 830 00 09, mobile phone 605 053 922 |
|
|
QUALITY MANAGER Marek Wróbel |
tel./fax 046 830 00 09, mobile phone 605 063 922 |
|
|
| |
|
The company is a member of the Polish Testing Laboratories Club POLLAB
|
|
|
The company cooperates with ANMAR Metrology Laboratory, Inc. from San Diego in California - USA
|
|
|